LETTER TO THE EDITOR
Year : 2022  |  Volume : 12  |  Issue : 3  |  Page : 113-114

Spuriously low end tidal carbon dioxide in capnometry: Nafion tube malfunction in end tidal carbon dioxide module blamed for near mishap!


1 Department of Anaesthesiology and Critical Care, All India Institute of Medical Sciences, Bhubaneswar, India
2 Department of Trauma and Emergency, All India Institute of Medical Sciences, Bhubaneswar, India
3 Department of Orthopedics, All India Institute of Medical Sciences, Bhubaneswar, India

Date of Submission03-Dec-2020
Date of Decision25-Jan-2021
Date of Acceptance28-Feb-2021
Date of Web Publication29-Nov-2021

Correspondence Address:
Chitta Ranjan Mohanty
Department of Trauma and Emergency, All India Institute of Medical Sciences, Bhubaneswar
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2045-9912.330695

Rights and Permissions

How to cite this article:
Ahmad SR, Mohanty CR, Bellapukonda S, Patro BP. Spuriously low end tidal carbon dioxide in capnometry: Nafion tube malfunction in end tidal carbon dioxide module blamed for near mishap!. Med Gas Res 2022;12:113-4

How to cite this URL:
Ahmad SR, Mohanty CR, Bellapukonda S, Patro BP. Spuriously low end tidal carbon dioxide in capnometry: Nafion tube malfunction in end tidal carbon dioxide module blamed for near mishap!. Med Gas Res [serial online] 2022 [cited 2022 Jan 22];12:113-4. Available from: https://www.medgasres.com/text.asp?2022/12/3/113/330695



Dear Editor,

We often fail to understand the technical complexity and related malfunction of newer sophisticated anesthesia workstations when encountered. Over the years there are many monitoring devices evolved for patient safety but malfunctioning of such devices gives a false sense of security and may adversely affect patient safety.[1] Anesthesia is like flying an aircraft, and most mishaps are during takeoff and landing. Like the flight data recorder, the anesthesia workstation records important information which can be analyzed later to find out the technical fault to ensure further patient safety.

A 38-year-old male patient was planned for clipping of middle cerebral artery aneurysm under general anesthesia. The pre-induction vitals were blood pressure 144/96 mmHg, heart rate 88 beats per minute. After uneventful induction, the patient was intubated and put-on mechanical ventilation. The ventilatory parameters in the workstation (GE Datex Ohmeda, Avance CS2’, Wipro GE Healthcare Pvt Ltd., Kolkota, West Bengal, India) were tidal volume of 500 mL, respiratory rate of 14/min, inspiratory to expiratory ratio of 1:2, and fractional inspired oxygen of 50%. The anesthesiologist noticed an initial end tidal carbon dioxide (ETCO2) of 17 mmHg on the patient monitor [Figure 1]. The tidal volume was reduced to 450 mL and respiratory rate to 12, but these changes in ventilatory parameters failed to correct the ETCO2. Few minutes later, the patient had profuse sweating and tachycardia. Arterial blood gas analysis revealed pH 7.22, partial pressure of carbon dioxide (PaCO2) 56 mmHg, partial pressure of oxygen 117 mmHg, bicarbonate 23.4 mM. A gradient between PaCO2 and ETCO2 of 39 seemed inappropriate. As PaCO2 was 56 mmHg, we increased the tidal volume to 550 mL and respiratory rate to 16/min. The repeat Arterial blood gas analysis displayed pH 7.40, PaCO2 34.3 mmHg, partial pressure of oxygen 128 mmHg, HCO3 23.7 mM. There was no sweating and heart rate returned to baseline value. The potential technical sites causing an erroneous ETCO2 value were analyzed as (1) external sampling tube (2) D-fend and (3) capnometer module. The D-fend is present in front of Datex-Ohmeda capnographs. It is a disposable water trap and filter that prevents moisture or humidity from entering the monitor. The external tube and D-fend were changed with no favorable result. So, we suspected a malfunction in the ETCO2 module or workstation. The intraoperative ventilatory parameters were monitored by repeat arterial blood gas analysis from arterial cannulation site. Rest of the perioperative course was uneventful. At the end of the surgery the patient was sent to intensive care unit where the patient was extubated later. A complaint was logged with GE Health Care (GE Healthcare Pvt Ltd., Kolkota, West Bengal, India) for the technical evaluation. A technical expert from the GE Health Care retrieved the intraoperative events data and we found that the ETCO2 was low from the initiation of mechanical ventilation. The technical expert thoroughly evaluated all the components and found the defective nafion tube inside the module (Carescape Respiratory Module E-sCAiO) was the cause of false low ETCO2 [Figure 1].
Figure 1: Ventilator control panel and display depicting the capnography with low end tidal carbon dioxide.
Note: Nafion tube (double red arrows) in the respiratory module.


Click here to view


Capnography works on the principal that carbon-dioxide absorbs infrared rays and the principle of Beer-Lambert law.[2] Infrared waves of specific wavelengths passing across the sampling gas are absorbed by the carbon-dioxide in it which is directly proportional to ETCO2. The remaining is detected by a sensor. The sampling tube in the respiratory module is made of nafion. Nafion is a sulfonated tetrafluoroethylene-based polymer highly permeable to water.[2] It equilibrates the gas for analysis with ambient air and removes excess water vapor if present. In our case, a defective nafion tube led to excessive clogging probably allowing insufficient gas to pass, resulting in erroneously low ETCO2.[3]

ETCO2 is a mandatory safety parameter in the perioperative period. In neurosurgical patients the ETCO2 must be monitored for maintenance of cerebral autoregulation[4]. Both hypercapnia as well as hypocapnia correlates with worse outcome. Hypocapnia as in our case alarmed us as it results in vasoconstriction decreasing cerebral blood flow with a resultant risk of cerebral ischemia.[4] This might have an additive effect on vasospasm induced decreased cerebral blood flow caused by subarachnoid haemorrhage. Low ETCO2 values may be caused by states of low cardiac output, hyperventilation, and pulmonary embolism. However, when pathologies are ruled out, technical glitches like damage to capnometry module, sampling line breakage, temperature or pressure effects and fault in ventilator itself must be looked into.[5] [Figure 2] shows step wise approach to low ETCO2 value in patient capnometry.[3] The abnormal ETCO2 values should be clinically correlated, and blood gas analysis must be done. Regular periodic maintenance of anesthesia workstations is essential to prevent this type of malfunction. We need to acquire some knowledge regarding the technical details of the machines used on a regular basis rather than waiting for a mishap to happen.
Figure 2: Flow diagram showing step wise approach to a low ETCO2 value in capnometry.
Note: ETCO2: End tidal carbon dioxide.


Click here to view


Written informed consent was obtained from the patient.

Open access statement: This is an open access journal, and articles are distributed under the terms of the Creative Commons AttributionNon- Commercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. ©Article authors (unless otherwise stated in the text of the article) 2022. All rights reserved. No commercial use is permitted unless otherwise expressly granted.



 
  References Top

1.
Mohanty CR, Ahmad SR. Drop in oxygen supply pressures: Misleading gas cylinder color coding to be blamed for nearing a mishap? J Clin Anesth. 2018;50:46-47.  Back to cited text no. 1
    
2.
Bhavani-Shankar K, Moseley H, Kumar AY, Delph Y. Capnometry and anaesthesia. Can J Anaesth. 1992;39:617-632.  Back to cited text no. 2
    
3.
Nonin Medical Inc. Operator’s Manual Lifesense® LS1-9R Vital Signs Monitor Capnography/Pulse Oximeter. https://www.nonin.com/wp- content/uploads/Operators-Manual-LifeSense.pdf. Accessed by June 15, 2021.  Back to cited text no. 3
    
4.
Deng RM, Liu YC, Li JQ, Xu JG, Chen G. The role of carbon dioxide in acute brain injury. Med Gas Res. 2020;10:81-84.  Back to cited text no. 4
    
5.
Fletcher R, Werner O, Nordström L, Jonson B. Sources of error and their correction in the measurement of carbon dioxide elimination using the Siemens-Elema CO2 Analyzer. Br J Anaesth. 1983;55:177-185.  Back to cited text no. 5
    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
References
Article Figures

 Article Access Statistics
    Viewed369    
    Printed8    
    Emailed0    
    PDF Downloaded40    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]